The FC1500 ensures a consistent charge power for medical laser and Intense Pulsed Light (IPL) therapy equipment, eliminating the necessity for multiple power supplies. By delivering a steady 1500 W of power across charge voltages ranging from 200 V to 1000 V, it significantly reduces treatment times. This constant power feature facilitates rapid capacitor charging at lower voltages in mixed technology systems.
The unit can be used as a stand alone capacitor charger and, if needed various low voltage modules can be added to the unit to provide further system rails.
Remote control and monitoring capability can be gained with the analogue and digital PMBus interfaces.
Module Name | Vnom (V) | Set point adjust range (V) | Imax (A) | Power (W) | Module Type |
---|---|---|---|---|---|
CmA | 5 | 2.5 - 6.0 | 21.0 | 105 | Single output (1 slot) |
CmB | 12 | 6.0 - 15.0 | 15.0 | 180 | Single output (1 slot) |
CmC | 24 | 15.0 - 28.0 | 8.3 | 200 | Single output (1 slot) |
CmD | 48 | 28.0 - 58.0 | 4.16 | 200 | Single output (1 slot) |
CmG V1 | 24 | 3.0 - 30.0 | 3.0 | 90 | Dual output (1 slot) |
CmG V2 | 24 | 3.0 - 30.0 | 3.0 | 90 | Dual output (1 slot) |
CmH V1 | 5 | 3.0 - 6.0 | 6.0 | 36 | Dual output (1 slot) |
CmH V2 | 24 | 3.0 - 30.0 | 3.0 | 90 | Dual output (1 slot) |
CmA-W01 | 5 | 1.0 - 6.0 | 21.0 | 105 | Wide trim (1 slot) |
CmB-W01 | 12 | 1.0 - 15.0 | 15.0 | 180 | Wide trim (1 slot) |
CmC-W01 | 24 | 2.0 - 28.0 | 8.3 | 200 | Wide trim (1 slot) |
CmD-W01 | 48 | 3.0 - 58.0 | 4.16 | 200 | Wide trim (1 slot) |
CmK | 200 | 175 - 205 | 0.66 | 132 | High voltage (1 slot) |
Up to 5 of any of these module options will fit in the FlexiCharge unit.
Product Highlights
Capacitor Charger:
- Up to 1500 Watts
- 0 to 1000 VDC
- Excellent pulse to pulse stability
- User power limit control
- Internal fan cooled
AC/DC Power Supply:
- Up to 800 Watts
- Up to 10 Isolated outputs
- Series and parallel outputs – higher voltages / currents
Typical Applications
Capacitor Charger:
- Medical laser application examples – Surgical lasers, laser treatment, laser aesthetic treatments
AC/DC Power Supply:
- Low voltage system power
- Power for onboard computers and displays
- Pump power
- Power for cooling and heating